A one-page read takes a manufacturer's public record (a 510(k) summary, a warning letter, a product
page) and reads it against the text that applies to it. It ends in the questions a reviewer would ask.
It is the first thing we send when someone asks what our work looks like.
These three follow real public records. Company names, product names and file numbers are removed,
and the record side is paraphrased so it does not point at anyone. The regulatory side is quoted word
for word, with its source, so you can check every line.
Sample 1 · FDA 524B · premarket
A desktop product moved to the web, and a warning letter followed
What the public record says
- An FDA warning letter issued in 2026 says the manufacturer moved a cleared software device from a desktop application to a web application.
- The letter says the change adds network-exposed attack surface and authentication risk, and that it requires a new premarket notification.
- It notes that FDA has not reviewed the device's SBOM or the documentation of third-party components affected by the migration.
What that triggers
whether intentionally or unintentionally, through any means
FDA guidance, Cybersecurity in Medical Devices (Feb 3, 2026), Section VII.B, on the “ability to connect to the internet”A web version meets that criterion, so the new 510(k) is a cyber device submission under section 524B.
a plan to monitor, identify, and address, as appropriate, in a reasonable time, postmarket cybersecurity vulnerabilities and exploits, including coordinated vulnerability disclosure and related procedures
Section 524B(b)(1), FD&C ActRequired in the submission itself, together with the SBOM under 524B(b)(3).
the documentation elements for the system threat modeling, cybersecurity risk assessment, Software Bill of Materials (SBOM), component support information, vulnerability assessments, and unresolved anomaly assessment(s)
FDA guidance, Cybersecurity in Medical Devices (Feb 3, 2026), Section V.A, contents of the security risk management report
Three questions a reviewer would ask
- Does the threat model cover the web tier (authentication, session handling, hosting environment), or was it written for the original desktop application?
- Does the SBOM list the third-party components of the web stack, each with its support status? That is the item the letter says FDA has not reviewed.
- Is the 524B(b)(1) plan written for the web product, with a coordinated vulnerability disclosure route a researcher can actually use?
Where to look first
Provide traceability between the threat model, cybersecurity risk assessment, SBOM, and testing documentation
FDA guidance, Cybersecurity in Medical Devices (Feb 3, 2026), Section V.AThe documents have to point to each other. A threat that never reaches the risk assessment, or an SBOM component no test touched, is a traceability finding even when every document exists.
- A desktop-era risk file carried over to a web product keeps the old attack surface and misses the new one the letter names.
Sample 2 · FDA 524B · postmarket
A gateway between the hospital PACS and the cloud
What the public record says
- A 510(k) summary for an imaging software device cleared in 2026 describes a gateway component that acts as a PACS listener: it receives a copy of each study, de-identifies it and transmits it to the manufacturer's cloud, where the analysis runs.
- When a result comes back, the gateway re-associates the preview image with the patient and routes it to the PACS.
- The summary states that a cybersecurity risk assessment and testing were conducted under FDA's guidance. It does not say where the gateway runs.
What that triggers
design, develop, and maintain processes and procedures to provide a reasonable assurance that the device and related systems are cybersecure
Section 524B(b)(2), FD&C Act, as quoted in the FDA guidance, Cybersecurity in Medical Devices (Feb 3, 2026)
software/firmware update servers, and connections to healthcare facility networks
FDA guidance, Cybersecurity in Medical Devices (Feb 3, 2026), Section VII.C, list of what FDA considers related systemsThe gateway, the cloud service and the hospital network connection all fall inside that list.
a plan to monitor, identify, and address, as appropriate, in a reasonable time, postmarket cybersecurity vulnerabilities and exploits, including coordinated vulnerability disclosure and related procedures
Section 524B(b)(1), FD&C ActIf the gateway runs on the hospital's side, the plan has to work for software on a network the manufacturer does not administer.
Three questions a reviewer would ask
- Where does the gateway run, and who patches it? On the hospital's side, an update has to pass the hospital's change control first.
- Does the threat model treat the gateway as a listener that accepts connections from the PACS and writes back to it, or only as a client of the cloud?
- Does the SBOM cover the gateway's own components (operating system, DICOM library, runtime), or only the cloud service?
Where to look first
The software component’s end-of-support date
FDA guidance, Cybersecurity in Medical Devices (Feb 3, 2026), Section V.A, SBOM elementsFor a gateway, that includes the operating system it runs on.
- A vulnerability plan that assumes the manufacturer can deploy a fix needs a second path for software it cannot reach without the customer.
Sample 3 · EU CRA · Article 14 reporting
A wellness wearable sold in the EU
What the public record says
- A wearable with a companion phone app is sold to consumers in the EU.
- The manufacturer's own terms say the product is not intended to diagnose or treat any condition.
- It was placed on the EU market before the Cyber Resilience Act applies in full.
What that triggers
This Regulation does not apply to products with digital elements to which the following Union legal acts apply: (a) Regulation (EU) 2017/745; (b) Regulation (EU) 2017/746
Regulation (EU) 2024/2847, Article 2(2)Those are the MDR and the IVDR. If the product is not a medical device under them, the exclusion does not reach it and the CRA applies.
Article 14 shall apply from 11 September 2026
Regulation (EU) 2024/2847, Article 71(2)
the obligations laid down in Article 14 shall apply to all products with digital elements that fall within the scope of this Regulation that have been placed on the market before 11 December 2027
Regulation (EU) 2024/2847, Article 69(3)A product already on the market owes the reporting duty today, before any other CRA requirement applies to it.
an early warning notification of an actively exploited vulnerability, without undue delay and in any event within 24 hours of the manufacturer becoming aware of it
Regulation (EU) 2024/2847, Article 14(2)(a)
Three questions a reviewer would ask
- Who decides that a vulnerability is actively exploited, and on what evidence?
- What starts the 24-hour clock in your process: the support ticket, the engineer's confirmation, or the decision of a named person?
- Which CSIRT is your coordinator, and does someone have working access to the single reporting platform?
Where to look first
a vulnerability for which there is reliable evidence that a malicious actor has exploited it in a system without permission of the system owner
Regulation (EU) 2024/2847, Article 3(42)The definition turns on evidence, so the process needs a named person who weighs it and a record of when the company became aware.
a software or hardware product and its remote data processing solutions
Regulation (EU) 2024/2847, Article 3(1)The app is software in its own right, and the cloud service the manufacturer runs for it can fall inside the product as remote data processing.
Want one for your device?
Request a one-page read: tell us what to read (a 510(k) number, a warning
letter, or a product page) and you get one page back by email. Free. No call.
If you already have a document written, the paid version of this is the
Document Review at $1,500 for FDA 524B, or the
EU In-Force-Today Pack at US$2,500 for CRA reporting.
Read against public documents only. Not legal advice. FDA guidance is nonbinding;
the statute is not. Sources: 21 U.S.C. 360n-2 (section 524B of the FD&C Act); FDA, Cybersecurity in
Medical Devices: Quality Management System Considerations and Content of Premarket Submissions
(February 3, 2026); Regulation (EU) 2024/2847 (Cyber Resilience Act).